There is growing interest in quantifying microplastics in environmental samples.
This application note presents a robust pressurized liquid extraction (PLE) with
pyrolysis-gas chromatography-mass spectrometry (pyr-GC/MS) method for
quantitation of microplastics like polyethylene (PE), polypropylene (PP), and
polystyrene (PS) at low concentrations in environmental matrices using the Agilent
5977B GC/MSD, Agilent 7890B GC, and Agilent MassHunter workstation software.
Linearity, limits of quantitation (LOQs), and reproducibility for real environmental
samples were evaluated. The GC/MSD addressed the insufficient limits of detection
that have challenged previous methods. PE, PP, and PS microplastics were
quantified down to 0.005 mg/g. Excellent linearity (R2
>0.97) for calibration samples
from 0.005 to 1 mg/g was obtained. Relative standard deviations (RSDs) for both
spiked and environmental samples were <10% or lower, demonstrating excellent
system reproducibility and reliability.
Za odpiranje dokumenta
kliknite tukaj …
01 4277 560- info@chemass.si
Promocije
Novice
GC and GC/MS Frequently Asked QuestionsObjavljeno: 13. 12. 2024From start to finish – Agilent CQA Analysis Solutions for IVT mRNA-based BiotherapeuticsObjavljeno: 4. 12. 2024Peak Scientific Product CatalogueObjavljeno: 6. 9. 2024Agilent Glycan Peak Assignment Tool (GPAT)Objavljeno: 19. 7. 2024Dogodki
Aplikacije
Glycan Peak Assignment Tool (GPAT)Objavljeno: 19. 7. 2024Quantification of Microplastics in Environmental Samples Using Pyrolysis and GC/MSDObjavljeno: 18. 11. 2020Fully Automated Characterization of Monoclonal Antibody Charge Variants Using 4D-LC/MSObjavljeno: 18. 11. 2020Accelerated Determination of Microplastics in Environmental Samples using TED-GC/MSObjavljeno: 18. 11. 2020



Ne zamudite naših prihodnjih dogodkov in obvestil
Digitalna preobrazba podjetja Chemass
Discuss with Agilent